IPL-808-200Jlaser energy meter specification
1. Optical energy measuring range: 1-100J, 200J
2. The detector's receiving aperture: 65x25mm
3. Spectral range: 350nm - 2500nm
4. Measurement uncertainty:<+-5%
Usage:
1. Insert the other end of the signal wire BNC provided with the probe to the right two holes of the digital multimeter (The protrusion is inserted upwards).
2. Turn the measurement gear from OFF to DC 600m
3. After the display data is stable, press the MAX MIN button three times, then the multimeter shows MAX-MIN in the upper left corner.
4. Trigger the light pulse, and the multimeter displays the monotonically rising data After 5 seconds,the display number shows a peak value and is held by the multimeter.
5. At this point, the value displayed on the multimeter is the difference between the maximum and the minimum which is the measured energy value(mV corresponds to J).
6. After measuring, press the MAX MIN button for 3 seconds to make the multimeter out of the state of MAXMIN. Observe the multimeter data, and repeat 3 to 5 steps to measure again after it stably displays and no longer drops monotonously.Light energy measurement range 1J-200J Effective receiving aperture of the detector 65mm * 25mm Spectral range 350nm-2500nm Measurement uncertainty 5%
-